CVE-2020-27278
What it is
In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, hard-coded credentials in the ventilator allow attackers with physical access to obtain admin privileges for the device’s configuration interface.
At a glance
- Exploitability
- Can be exploited with physical access without logging in. High risk of data exposure.
- Exploit odds
- 0.3% chance of exploitation in the next 30 days (EPSS).
- Fix
- No patch or advisory is linked yet. Look at mitigations or restrict exposure.
Timeline
- PublishedDisclosed publicly on CVE.org and NVD
- Last updatedMost recent change to the record
Why it ranks here
- Exploited in the wild: no, CISA does not list it.
- Exploit odds: 0.3% in the next 30 days (EPSS).
- Severity: 5.2 MEDIUM (CVSS).
- Fix: none linked yet, which raises it.
Details
- Published
- Last modified
- Exploit probability
- 0.3% · EPSS, next 30 days
- Weakness
- CWE-798 · Hard-coded credentials
- Record status
- Modified
Affected
- hamilton-medical hamilton-t1 Firmware ≤ 2.2.3
Attack profile CVSS 3.1 · 5.2
- Attack vector
- Physical
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- High
- Integrity
- Low
- Availability
- None
Raw vector
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:N