CVE-2018-8870
What it is
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.
At a glance
- Exploitability
- Can be exploited with physical access without logging in. High risk of data exposure, data tampering, service disruption.
- Exploit odds
- 0.4% chance of exploitation in the next 30 days (EPSS).
- Fix
- No patch or advisory is linked yet. Look at mitigations or restrict exposure.
Timeline
- PublishedDisclosed publicly on CVE.org and NVD
- Last updatedMost recent change to the record
Why it ranks here
- Exploited in the wild: no, CISA does not list it.
- Exploit odds: 0.4% in the next 30 days (EPSS).
- Severity: 6.4 MEDIUM (CVSS).
- Fix: none linked yet, which raises it.
Details
- Published
- Last modified
- Exploit probability
- 0.4% · EPSS, next 30 days
- Weakness
- CWE-259
- Record status
- Modified
Affected
- Medtronic 24950 Mycarelink Monitor Firmware
- Medtronic 24952 Mycarelink Monitor Firmware
Attack profile CVSS 3.1 · 6.4
- Attack vector
- Physical
- Complexity
- High
- Privileges needed
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Raw vector
CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H