CVE-2018-18562
What it is
An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before 03.01.04 and CoaguChek / cobas h232 Handheld Base Unit before 03.01.04. Weak access credentials may enable attackers in the adjacent network to gain unauthorized service access via a service interface.
At a glance
- Exploitability
- Can be exploited from the same network segment without logging in. High risk of data exposure, data tampering, service disruption.
- Exploit odds
- 0.7% chance of exploitation in the next 30 days (EPSS).
- Fix
- A patch, advisory or fixed version is referenced. Upgrade or apply it.
Timeline
- PublishedDisclosed publicly on CVE.org and NVD
- Last updatedMost recent change to the record
Why it ranks here
- Exploited in the wild: no, CISA does not list it.
- Exploit odds: 0.7% in the next 30 days (EPSS).
- Severity: 8.8 HIGH (CVSS).
- Fix: a patch or advisory is linked.
Details
- Published
- Last modified
- Exploit probability
- 0.7% · EPSS, next 30 days
- Weakness
- CWE-521
- Record status
- Modified
Affected
- Roche accu-chek Inform Ii Firmware < 03.01.04
- Roche Cobas H 232 Firmware < 03.01.04
- Roche Coaguchek Firmware < 03.01.04
- Roche Base Unit Hub Firmware < 03.01.04
Attack profile CVSS 3.0 · 8.8
- Attack vector
- Adjacent network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Raw vector
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Is it in your code?
Hand this to your coding agent. It searches your project for the affected products and checks versions. Nothing leaves your machine.
Read https://cve-list.com/agent/cve/CVE-2018-18562.md and follow it. Check whether this project is affected by CVE-2018-18562. Read only. Don't change anything and don't upload my code. Tell me yes, no or not sure, with the file as proof.
Fix and advisories
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-310-01
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-310-01