CVE-2017-9657
What it is
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
At a glance
- Exploitability
- Can be exploited from the same network segment without logging in. High risk of service disruption.
- Exploit odds
- 0.8% chance of exploitation in the next 30 days (EPSS).
- Fix
- A patch, advisory or fixed version is referenced. Upgrade or apply it.
Timeline
- PublishedDisclosed publicly on CVE.org and NVD
- Last updatedMost recent change to the record
Why it ranks here
- Exploited in the wild: no, CISA does not list it.
- Exploit odds: 0.8% in the next 30 days (EPSS).
- Severity: 6.5 MEDIUM (CVSS).
- Fix: a patch or advisory is linked.
Details
- Published
- Last modified
- Exploit probability
- 0.8% · EPSS, next 30 days
- Weakness
- CWE-460
- Record status
- Modified
Affected
- Philips Intellivue mx40 Firmware < b.06.18
Attack profile CVSS 3.0 · 6.5
- Attack vector
- Adjacent network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- None
- Integrity
- None
- Availability
- High
Raw vector
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Is it in your code?
Hand this to your coding agent. It searches your project for the affected products and checks versions. Nothing leaves your machine.
Read https://cve-list.com/agent/cve/CVE-2017-9657.md and follow it. Check whether this project is affected by CVE-2017-9657. Read only. Don't change anything and don't upload my code. Tell me yes, no or not sure, with the file as proof.
Fix and advisories
- https://www.usa.philips.com/healthcare/about/customer-support/product-security
- https://www.usa.philips.com/healthcare/about/customer-support/product-security
References
- http://www.securityfocus.com/bid/100813
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01
- https://www.usa.philips.com/healthcare/about/customer-support/product-security
- http://www.securityfocus.com/bid/100813
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01
- https://www.usa.philips.com/healthcare/about/customer-support/product-security