CVE-2017-12725
What it is
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
At a glance
- Exploitability
- Can be exploited over the network without logging in. Limited direct impact.
- Exploit odds
- 0.8% chance of exploitation in the next 30 days (EPSS).
- Fix
- No patch or advisory is linked yet. Look at mitigations or restrict exposure.
Timeline
- PublishedDisclosed publicly on CVE.org and NVD
- Last updatedMost recent change to the record
Why it ranks here
- Exploited in the wild: no, CISA does not list it.
- Exploit odds: 0.8% in the next 30 days (EPSS).
- Severity: 5.6 MEDIUM (CVSS).
- Fix: none linked yet, which raises it.
Details
- Published
- Last modified
- Exploit probability
- 0.8% · EPSS, next 30 days
- Weakness
- CWE-798 · Hard-coded credentials
- Record status
- Modified
Affected
- smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1
- smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump 1.5
- smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump 1.6
Attack profile CVSS 3.0 · 5.6
- Attack vector
- Network
- Complexity
- High
- Privileges needed
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- Low
- Integrity
- Low
- Availability
- Low
Raw vector
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L