CVE-2015-3956
What it is
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
At a glance
- Exploitability
- Can be exploited over the network without logging in. High risk of data exposure, data tampering, service disruption.
- Exploit odds
- 1.0% chance of exploitation in the next 30 days (EPSS).
- Fix
- A patch, advisory or fixed version is referenced. Upgrade or apply it.
Timeline
- PublishedDisclosed publicly on CVE.org and NVD
- Last updatedMost recent change to the record
Why it ranks here
- Exploited in the wild: no, CISA does not list it.
- Exploit odds: 1.0% in the next 30 days (EPSS).
- Severity: 9.8 CRITICAL (CVSS).
- Fix: a patch or advisory is linked.
Details
- Published
- Last modified
- Exploit probability
- 1.0% · EPSS, next 30 days
- Weakness
- CWE-345
- Record status
- Modified
Affected
- Pifzer Plum a\+ Infusion System Firmware ≤ 13.4
- Pifzer Plum a\+3 Infusion System Firmware ≤ 13.6
- Pifzer Symbiq Infusion System Firmware ≤ 3.13
Attack profile CVSS 3.0 · 9.8
- Attack vector
- Network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Raw vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Is it in your code?
Hand this to your coding agent. It searches your project for the affected products and checks versions. Nothing leaves your machine.
Read https://cve-list.com/agent/cve/CVE-2015-3956.md and follow it. Check whether this project is affected by CVE-2015-3956. Read only. Don't change anything and don't upload my code. Tell me yes, no or not sure, with the file as proof.
Fix and advisories
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01