CVE-2015-3459
What it is
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
At a glance
- Exploitability
- Can be exploited over the network without logging in. High risk of data exposure, data tampering, service disruption.
- Exploit odds
- 5.1% chance of exploitation in the next 30 days (EPSS).
- Fix
- No patch or advisory is linked yet. Look at mitigations or restrict exposure.
Timeline
- ReservedThe CVE id was reserved by its CNA
- PublishedDisclosed publicly on CVE.org and NVD
- Last updatedMost recent change to the record
Why it ranks here
- Exploited in the wild: no, CISA does not list it.
- Exploit odds: 5.1% in the next 30 days (EPSS).
- Severity: 10.0 HIGH (CVSS).
- Fix: none linked yet, which raises it.
Details
- Published
- Last modified
- Exploit probability
- 5.1% · EPSS, next 30 days
- Weakness
- CWE-264
- Record status
- Modified
Affected
- Hospira Lifecare Pcainfusion Firmware ≤ 5.0
- Hospira Lifecare pca3
- Hospira Lifecare pca5
Attack profile CVSS 2.0 · 10.0
- Attack vector
- Network
- Complexity
- Low
- Authentication
- None
- Confidentiality
- Complete
- Integrity
- Complete
- Availability
- Complete
Raw vector
AV:N/AC:L/Au:N/C:C/I:C/A:C
Is it in your code?
Hand this to your coding agent. It searches your project for the affected products and checks versions. Nothing leaves your machine.
Read https://cve-list.com/agent/cve/CVE-2015-3459.md and follow it. Check whether this project is affected by CVE-2015-3459. Read only. Don't change anything and don't upload my code. Tell me yes, no or not sure, with the file as proof.
References
- http://hextechsecurity.com/?p=123
- http://imgur.com/CEAnZjj
- http://imgur.com/JHiWSqd
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- http://www.securityfocus.com/bid/74414
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
- https://twitter.com/dyngnosis/status/592671049487142913
- https://twitter.com/dyngnosis/status/592743461977219072
- http://hextechsecurity.com/?p=123
- http://imgur.com/CEAnZjj
- http://imgur.com/JHiWSqd
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- http://www.securityfocus.com/bid/74414
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
- https://twitter.com/dyngnosis/status/592671049487142913
- https://twitter.com/dyngnosis/status/592743461977219072