CVE-2013-2808
What it is
Heap-based buffer overflow in Xper in Philips Xper Information Management Physiomonitoring 5 components, Xper Information Management Vascular Monitoring 5 components, and Xper Information Management servers and workstations for Flex Cardio products before XperConnect 1.5.4.053 SP2 allows remote attackers to execute arbitrary code via a crafted HTTP request to the Connect broker on TCP port 6000.
At a glance
- Exploitability
- Can be exploited over the network without logging in. High risk of data exposure, data tampering, service disruption.
- Exploit odds
- 4.3% chance of exploitation in the next 30 days (EPSS).
- Fix
- No patch or advisory is linked yet. Look at mitigations or restrict exposure.
Timeline
- ReservedThe CVE id was reserved by its CNA
- PublishedDisclosed publicly on CVE.org and NVD
- Last updatedMost recent change to the record
Why it ranks here
- Exploited in the wild: no, CISA does not list it.
- Exploit odds: 4.3% in the next 30 days (EPSS).
- Severity: 9.3 HIGH (CVSS).
- Fix: none linked yet, which raises it.
Details
- Published
- Last modified
- Exploit probability
- 4.3% · EPSS, next 30 days
- Weakness
- CWE-119 · Buffer errors
- Record status
- Modified
Affected
- Philips Xper Information Management Physiomonitoring 5
- Philips Xperconnect ≤ 1.5.4.053
- Philips Xper Information Management Vascular Monitoring 5
- Philips Xper Flex Cardio
Attack profile CVSS 2.0 · 9.3
- Attack vector
- Network
- Complexity
- Medium
- Authentication
- None
- Confidentiality
- Complete
- Integrity
- Complete
- Availability
- Complete
Raw vector
AV:N/AC:M/Au:N/C:C/I:C/A:C